Specifications

Gases Measured

Carbon Dioxide (CO2)

Method of Detection

Infrared detector

Ranges Available

0-20% CO2

Resolution

0.1% on %, 1 PPM on PPM

Accuracy & Repeatability

± 1-2% of full scale

Drift

Less than 2% of full scale per month

Response Time (T-90)

20-30 seconds to 90% of step change

Ambient Temperature Range

32 to 122°F (0 to 50°C) @ 5-95% RH non-condensing. Lower temperatures (-5°F, -20°C) with Cold Weather Package. Sample dew point should be at least 9°F (5°C) below lowest sample temperature.

Linearity

±1% of full scale

Size & Weight

Dimensions and weight will vary with configurations selected

Power

115 VAC, 60 Hz
220 VAC, 50 Hz

Alarms

High or low alarm contacts available for all gases and low flow

Code Numbers for Gases Measured

CO2 by infrared detector

Sampling System

Std SS probe with unheated pre-filter (Std. probe 3/8" O.D. x 18") (9.5 x 460 mm)

Display Type

D1 - LED display (s) - Manual potentiometer calibration
D2 - LED display (s) - Manual push-button calibration
D3 - Microprocessor controller with single graphic display. Allows auto sequencing, RS-485 output, 'CAL NOW' and one-step
D4 - Same as D3 but with full auto-calibration

Output Options

OP1 - 4-20 mA non-isolated
OP2 - 4-20 mA isolated
OP3 - 0-5 V
OP4 - 0-10 V
OP5 - RS-485
OP6 - Modbus

Analyzer Type

Flue Gas Analyzer

Industry Type

Flue Gas or Emissions


Applications
    Continuous analysis of flue gas for any combination and range of O2, CO, CO2, SO2, NOx (NO or NO + NO2) or combustibles from combustion processes.



Features
  • Best value continuous flue gas analyzer with modular sample conditioning
  • Oxygen detected by either long life electrochemical sensor, high accuracy paramagnetic sensor or zirconia sensor, depending on desired accuracy and sample composition
  • Fast response time and bright digital readout
  • Microprocessor based with push button calibration
  • Easy to maintain modular layout
  • Sample preparation includes moisture removal.



Options
  • Hi/Low gas, low flow, and diagnostic alarms available
  • Isolated analog, RS232, RS485, MODBUS, and Ethernet outputs available
  • Cabinet purge system available for use in hazardous areas
  • One-step 'cal-now' button with touch screen LCD display
  • Full automatic calibration with touch screen LCD display and low flow switch
  • Cold weather package for operation to -5°F (-20°C)
  • Cabinet coolers can be fitted to most models
  • Sensors mounted in heated box for reduced drift
  • High temperature filters and probes
Optional sample conditioning includes:
  1. Continuous water pre wash system for acidic fuel gases present but not to be analyzed (Model 7300)
  2. Nafion-Tube™ based sample dryer for high accuracy analysis of acidic samples, where NO2 or SO2 are to be measured (Model 7250)
  3. Heated stack filter for removal of dust from high dust load (> µ) samples (Model 7240)
  4. Calibration gas injected at heated stack filter to minimize discrepancies due to absorption of soluble gases.
  5. Multiple sample point sequencer which allows multiple samples to be selected and analyzed (Model 402AS)