Specifications

Method of Detection

Temperature-controlled TC cell for H2; separate NDIR detector for CO2

Ranges Available

0-10.0% to 0-100% H2; and 0-15.0% to 0-100% CO2

Resolution

0.1% H2 and CO2

Accuracy & Repeatability

±1% of full scale

Drift

Zero and Span ±1% of full scale per week

Response Time (T-90)

10-15 seconds

Ambient Temperature Range

32 to 122°F (0 to 50°C), Lower temperatures (-5°F, -20°C) with Cold Weather Package

Linearity

Within 1% of full scale

Size & Weight

Dimensions will vary depending on enclosure style and options required

Power

115 VAC, 60 Hz
220 VAC, 50 Hz

Output Options

4-20 mA into 500 ohms non-isolated standard
Isolated 4-20 mA, RS232, RS485, MODBUS, Ethernet outputs

Alarms

High and/or low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating
Low flow alarm

Analyzer Type

H2 Analyzer

Hazardous Area Rating Options

Class 1, Div 1, Grp A-D Hazardous Area Rating


Applications
    For continuous analysis of hydrogen (H2) with active carbon dioxide (CO2) compensation. Recommended for continuous analysis of exothermic (DX) atmosphere gases from a generator or a furnace, and other process gases.




Features
  • High accuracy - continuous CO2 compensation of H2 reading
  • Bright digital readout, and voltage or 4 - 20 mA outputs
  • Long life thermal conductivity (TC) detector with high accuracy and repeatability, cannot be burned out from loss of flow
  • Separate NDIR CO2 detector
  • Easy-to-maintain modular layout
  • Built in sample pump, filters, and flowmeter




Options
  • Hi/Low H2, CO2, low flow, and diagnostic alarms available
  • Isolated analog, RS232, RS485, MODBUS, and Ethernet outputs available
  • Sample conditioning systems available for removal of moisture, acidic gas, and dust
  • Cabinet purge system available for use in hazardous areas
  • One-step 'cal-now' button with touch screen LCD display
  • Full automatic calibration with touch screen LCD display
  • Cabinet coolers can be fitted to most models
  • Sensors mounted in temperature controlled heated box to improve accuracy
  • High temperature filters and probes




Calibration
  • On nitrogen or air for zero
  • On a blend of 80-90% of full scale range of H2 & CO2 for span