Specifications

Method of Detection

Temperature-controlled TC cell for H2; separate NDIR detector for CO2

Ranges Available

0-10.0% to 0-100% H2; and 0-15.0% to 0-100% CO2

Resolution

0.1% H2 and CO2

Accuracy & Repeatability

±1% of full scale

Drift

Zero and Span ±1% of full scale per week

Response Time (T-90)

10-15 seconds

Ambient Temperature Range

32 to 122°F (0 to 50°C), Lower temperatures (-5°F, -20°C) with Cold Weather Package

Linearity

Within 1% of full scale

Size & Weight

Dimensions will vary depending on enclosure style and options required

Power

115 VAC, 60 Hz
220 VAC, 50 Hz

Output Options

4-20 mA into 500 ohms non-isolated standard
Isolated 4-20 mA, RS232, RS485, MODBUS, Ethernet outputs

Alarms

High and/or low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating
Low flow alarm

Analyzer Type

H2 Analyzer


Applications
    For continuous analysis of hydrogen (H2) with active carbon dioxide (CO2) compensation. Recommended for continuous analysis of exothermic (DX) atmosphere gases from a generator or a furnace, and other process gases.




Features
  • High accuracy - continuous CO2 compensation of H2 reading
  • Bright digital readout, and voltage or 4 - 20 mA outputs
  • Long life thermal conductivity (TC) detector with high accuracy and repeatability, cannot be burned out from loss of flow
  • Separate NDIR CO2 detector
  • Easy-to-maintain modular layout
  • Built in sample pump, filters, and flowmeter




Options
  • Hi/Low H2, CO2, low flow, and diagnostic alarms available
  • Isolated analog, RS232, RS485, MODBUS, and Ethernet outputs available
  • Sample conditioning systems available for removal of moisture, acidic gas, and dust
  • Cabinet purge system available for use in hazardous areas
  • One-step 'cal-now' button with touch screen LCD display
  • Full automatic calibration with touch screen LCD display
  • Cabinet coolers can be fitted to most models
  • Sensors mounted in temperature controlled heated box to improve accuracy
  • High temperature filters and probes




Calibration
  • On nitrogen or air for zero
  • On a blend of 80-90% of full scale range of H2 & CO2 for span